• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 87035
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2022
Event Type  malfunction  
Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported that during a radiofrequency ablation procedure, an intellamap orion was selected for use.The irrigation of the catheter was blocked.When the pressure bag line was connected to the catheter, it did not start to flow.Trying to flush the orion with a syringe also was not possible.Exchanging the catheter resolved the issue, and the procedure was completed successfully.The device was disposed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key15633454
MDR Text Key302032391
Report Number2124215-2022-42234
Device Sequence Number1
Product Code DRF
UDI-Device Identifier08714729841968
UDI-Public08714729841968
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K143481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/18/2023
Device Model Number87035
Device Catalogue Number87035
Device Lot Number0028882566
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/13/2022
Initial Date FDA Received10/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/16/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-