• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN HOUSING RESECT COLLET S3-8; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN HOUSING RESECT COLLET S3-8; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71434424
Device Problems Fracture (1260); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2022
Event Type  malfunction  
Event Description
It was reported that, during a tka surgery, it was noticed that an lgn housing resect collet s3-8 had become stuck on the box chisel instrument, due to it splintering.The procedure was resumed, without any delay, changing the surgical technique.No other complications were reported.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
Section h10: the device was not returned for evaluation and the reported event could not be confirmed.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
 
Event Description
It was reported that, during a tka surgery, it was noticed that an lgn housing resect collet s3-8 had become stuck on the box chisel instrument, due to it splintering.The procedure was resumed, without any delay, by using a double action nibbler.Nothing fell in the patient.No other complications were reported.
 
Event Description
It was reported that, during a tka surgery, it was noticed that an lgn housing resect collet s3-8 had become stuck on the box chisel instrument, due to it splintering.The procedure was resumed, without any delay, by using a double action nibbler.Nothing fell in the patient.No other complications were reported.
 
Manufacturer Narrative
Additional information: b5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN HOUSING RESECT COLLET S3-8
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15636062
MDR Text Key306697671
Report Number1020279-2022-04489
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010549327
UDI-Public03596010549327
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71434424
Device Catalogue Number71434424
Device Lot Number09MM06315
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/28/2022
Initial Date FDA Received10/19/2022
Supplement Dates Manufacturer Received10/20/2022
11/16/2022
Supplement Dates FDA Received10/23/2022
11/21/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/10/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-