SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number C1120
Device Problems Crack (1135); Leak/Splash (1354); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the unit's inspiratory filter connection was cracked, there was a hole in circuit, and there was leaking when water test was performed.No patient injury was reported.
 
Manufacturer Narrative
(udi) and g5 (510k) are unknown; no additional information was provided.This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No lot number was provided; therefore, a device history record (dhr) review could not be completed.The sample consisted of three (3) products from unknown lots, received in used conditions without their original packaging.Visual and functional testing were performed.Visual inspection: the samples were visually inspected at a distance of 12 inches under normal lighting to received unit, in order to detect any damage on the unit; two (2) samples have a tear.Functional test: one (1) sample was tested due to confirmation of leaking failure mode.Leak test was performed and the circuit did not pass the leak test.The root cause was deemed to be manufacturing-related.Actions (s) taken to mitigate the reported issue(s): a corrective and preventative action has been opened to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15638194
MDR Text Key307164156
Report Number3012307300-2022-25352
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC1120
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/03/2022
Initial Date FDA Received10/20/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-