• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS Back to Search Results
Catalog Number 8606500
Device Problems Decrease in Pressure (1490); Inaccurate Delivery (2339); No Pressure (2994)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/04/2022
Event Type  malfunction  
Event Description
It was reported that duriing a case the airway pressure was lost.There was no injury reported.The device reportedly generated an alarm.
 
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event Description
It was reported that duriing a case the airway pressure was lost.There was no injury reported.The device reportedly generated an alarm.
 
Manufacturer Narrative
The electronic log file was available for investigation.The case in question could be reconstructed by means of the information stored therein.Beginning at 5:07pm minute volume leaks of up to 4.4 l/min in the patient circuit were present.The log entries do not indicate a device failure.The device detected the external leak and reacted as specified for such situation as it posted several alarms (e.G.Fresh gas low or leakage, apnea).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
MDR Report Key15645578
MDR Text Key303744533
Report Number9611500-2022-00284
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8606500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/20/2022
Initial Date FDA Received10/21/2022
Supplement Dates Manufacturer Received11/07/2022
Supplement Dates FDA Received12/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-