MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number C37101330-NLJ
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2019
Event Type  malfunction  
Manufacturer Narrative
(udi) and (510k) are unknown; no additional information was provided.This mdr was generated under protocol b10009704, as a result of warning letter cms#: 617147.No lot number was provided; therefore, a device history record (dhr) review could not be completed.The sample was received in used conditions without its original packaging.Visual inspection: the sample was visually inspected at a distance of 12 inches under normal lighting; no discrepancies detected.Functional test: a leak test was performed; the breathing circuit did not pass the test.The most probable root cause of the issue was deemed to be manufacturing-related.Action(s) taken to mitigate the reported issue(s): a corrective and preventative action has been opened to address the reported issue.
 
Event Description
It was reported that, during a pre-use check, it was noticed that air was leaking from either the anesthesia circuit or the bag.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
motohakma, naka-ku, yokohama
minneapolis, MN 55442
MDR Report Key15650429
MDR Text Key306933372
Report Number3012307300-2022-25680
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC37101330-NLJ
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/03/2022
Initial Date FDA Received10/21/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-