Catalog Number C49101320J
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/20/2018
Event Type  malfunction  
Manufacturer Narrative
(udi) and (510k) are unknown; no additional information was provided.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No lot number was provided; therefore, a device history record (dhr) review could not be completed.The sample was received in used conditions without its original packaging.Visual inspection: the sample was visually inspected at a distance of 12 inches under normal lighting; there was a tear on the tube.Functional test: a leak test was performed; the breathing circuit did not pass the test.The most probable root cause of the issue was deemed to be manufacturing-related.Action(s) taken to mitigate the reported issue(s): a corrective and preventative action has been opened to address the reported issue.
 
Event Description
It was reported that, after establishing a connection of the product and when performing a pre-use air leak check on it, air leaked from the part where a scotch tape was affixed.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15650456
MDR Text Key306893279
Report Number3012307300-2022-25682
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC49101320J
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/03/2022
Initial Date FDA Received10/21/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-