Model Number OER-PRO
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/26/2022
Event Type  malfunction  
Manufacturer Narrative
An olympus field service engineer (fse) was dispatched to evaluate and repair the device.The fse replaced the yellow scope connector in basin, tested and the unit was functional.The fse serviced the subject device, and it passed the electrical safety test.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus technical assistance center (tac), one of the channel connectors of the endoscope reprocessor was damaged.There was no patient involvement in this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to h3.Please see updates to h3, h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the investigation, it¿s likely the connector became loose and was broken.Therefore, the connecting tube was unable to be connected.It¿s probable the connector became loose due to accumulated stress or the user hit a hard object to the connector.However, a final root cause was unable to be identified.The event can be detected by following the instructions for use which state: ¿chapter 3.Inspection before use 3.3 inspecting the connectors check the following for each connector.-the connector should be fixed firmly.-the o-rings should be free of abnormalities such as cracks, tears, or dents.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15650804
MDR Text Key303643767
Report Number9610595-2022-03178
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/27/2022
Initial Date FDA Received10/21/2022
Supplement Dates Manufacturer Received11/28/2022
Supplement Dates FDA Received12/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/10/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-