• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number C37101307J
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2019
Event Type  malfunction  
Event Description
It was reported that, during a pre-use check, the customer found leakage of air around the connection part of the anesthesia circuit.The product was not used.No patient injury was reported.
 
Manufacturer Narrative
(udi) and g5 (510k) are unknown; no additional information was provided.This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.The sample was received in used conditions without its original packaging.Visual inspection: the sample was visually inspected at a distance of 12 inches under normal lighting; no discrepancies detected.Functional test: a leak test was performed, the circuit did not pass the leak test.The most probable root cause of the issue was deemed to be manufacturing related.Action(s) taken to mitigate the reported issue(s): a corrective and preventative action has been opened to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
chuoh-ku, osaka-city
minneapolis, MN 55442
MDR Report Key15651461
MDR Text Key306890536
Report Number3012307300-2022-25706
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/25/2022
Device Catalogue NumberC37101307J
Device Lot Number3849914
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/04/2022
Initial Date FDA Received10/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-