Updated fields: b4, d9, e1(site country), g3, g6, h2, h3, h4, h6(type of investigation, investigation findings, component codes, investigation conclusions), h10, h11.It was reported before use that the cs300 intra-aortic balloon pump (iabp) had maintenance code 4.There was no patient involvement and no harm reported.A getinge field service engineer (fse) evaluated and was not able to reproduce the error.Fan is functioning normally.Fse stated that unit might have been placed against corner.Customer left unit running for half day and error came back.Fse said that main board will need to be replaced but main board didn't resolve the issue.He found damage fan cable and replaced cable,main bd thermoswitch/fan and switch, thermal and error cleared.Performed functional testing and passed.Electrical safety passed.Unit return to the customer for clinical use.
|