Model Number CYF-VH
Device Problems Device Reprocessing Problem (1091); Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, a tear was discovered inside the channel and minor deformations were found in the video connector.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
The customer reported that her olympus cysto-nephro videoscope had debris found inside the scope during reprocessing.There was no patient involvement during this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the likely cause of the reported event is due to physical stress.However, the root cause of the reported event is unable to be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15657828
MDR Text Key307111550
Report Number3002808148-2022-03452
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170310461
UDI-Public04953170310461
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/29/2022
Initial Date FDA Received10/24/2022
Supplement Dates Manufacturer Received11/18/2022
Supplement Dates FDA Received12/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-