Model Number X |
Device Problems
Communication or Transmission Problem (2896); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/10/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the pressurewire x, wireless device was calibrated.The device was to be used in a lesion in the left anterior descending (lad) artery.However, after equalization was performed, the device lost signal during the procedure despite several attempts to reconnect the device.Therefore, the device was removed and the procedure was completed with another pressurewire x, wireless device.There were no adverse patient effects and no clinically significant delay in the procedure.The device return analysis revealed the distal tube material was sporadically torn and bunched 2.5cm distal to the proximal end for a length of 2.5cm.No additional information was provided.
|
|
Manufacturer Narrative
|
Visual analysis was performed on the returned device.The reported signal loss was unable to be confirmed due to the condition of the returned unit prevented functional testing.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no indication of a lot specific product quality issue.The investigation was unable to determine a cause for the reported communication or transmission problem (signal loss).It may be possible that damage caused by mechanical/external force caused damage to the microcables resulting in loss of signal; however, this could not be confirmed.Returned analysis noted movement of the sensor jacket, tears to the distal tube material, and stretched tip coils suggest that the distal end of the wire was likely damaged during removal, possibly due to interaction with the guide catheter upon retraction.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
Search Alerts/Recalls
|
|