Catalog Number 670012
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2018
Event Type  malfunction  
Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No product information has been provided to date.No information has been provided to date.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of product.A product sample was received for evaluation.Visual and functional testing were performed.One sample was received in used conditions with its opened original packing inside a ziploc bag.The sample was visually inspected with a 12 inches distance under normal conditions of illumination.During the visual inspection, it was seen that sample was broken in two parts.The root cause was unable to be determined, most likely the product was damaged after leaving the manufacturing facility.As a preventive action production personnel was notified by quality engineer as awareness of the defect reported by the customer.
 
Event Description
It was reported that during the pre-use air leak check on the product, the product failed.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA BAGS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15660422
MDR Text Key306516953
Report Number3012307300-2022-26060
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2020
Device Catalogue Number670012
Device Lot Number3475697
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/30/2022
Initial Date FDA Received10/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-