Model Number EG16-K10 |
Device Problems
Partial Blockage (1065); Collapse (1099); Failure to Disconnect (2541)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/19/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device is not distributed in the usa, therefore the pma/510(k) number is not applicable.If additional information becomes available, a supplemental report will be filed with the new information.
|
|
Event Description
|
A brush stuck in the op-channel this event occurred at the time of during reprocessing.There was no report of patient harm.
|
|
Manufacturer Narrative
|
Correction information g6: follow up #1 h2:if follow-up, what type? h3:device evaluated by manufacture h6: coding changed based on the investigation result additional information h4:device manufacture date.Evaluation summary due to a stuck cleaning brush due to using an incompatible cleaning brush.
|
|