Patient's machine started to alarm "flow balance inaccurate" and "flow balance error" during his treatment and we were unable to clear this alarm.Rinseback was given and patient chose to leave instead of wait for another machine to be set up.He had 1 hour and 26 minutes left.Biomed was notified.Pa was also notified.Manufacturer response for hemodialysis unit, (brand not provided) (per site reporter).Agreed with repair plan.The root cause of the failure was the p2 pump output was not within standards.When i would complete the p2 test the reading were all over the place until i replaced the pump.
|