Model Number STRIP, KETONE 50CT
Device Problem Unexpected Color (4055)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/02/2022
Event Type  malfunction  
Event Description
Consumer reported complaint for the ketone test strips.Complaint was reported via e-mail.Customer's e-mail stated that the ketone test strips "don¿t work at all" and that "the color just turns muddy looking and does not come close to anything on the color chart." no symptoms and no medical intervention related to the use of the product was reported.Customer did not provide contact telephone number; email response had been sent to customer.
 
Manufacturer Narrative
Internal report reference number: (b)(4).Ketone test strips were not returned for evaluation.Note: manufacturer contacted customer in follow-up e-mails to ensure the initial concern is resolved - unable to establish contact with customer at this time.
 
Manufacturer Narrative
Sections with additional information as of 22-nov-2022: h6: updated fda's type of investigation, investigation findings, and investigation conclusions.H10:ketone strips were not returned for evaluation.Retention testing was performed using test strips from the same lot.Retention strips tested within specifications.Most likely underlying root cause: mlc-020: user's test strip had poor storage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KETONE
Type of Device
NITROPRUSSIDE, KETONES (URINARY, NON-QUANT.)
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key15666279
MDR Text Key303239887
Report Number1000113657-2022-00569
Device Sequence Number1
Product Code JIN
UDI-Device Identifier00021292008178
UDI-Public021292008178
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/19/2023
Device Model NumberSTRIP, KETONE 50CT
Device Lot NumberAA657
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date10/02/2022
Initial Date Manufacturer Received 10/02/2022
Initial Date FDA Received10/25/2022
Supplement Dates Manufacturer Received11/01/2022
Supplement Dates FDA Received11/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-