• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S X-FLOW CATHETER; CATHETER, RETENTION TYPE, BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S X-FLOW CATHETER; CATHETER, RETENTION TYPE, BALLOON Back to Search Results
Model Number AB63221002
Device Problem Burst Container or Vessel (1074)
Patient Problem Hematuria (2558)
Event Date 09/14/2022
Event Type  malfunction  
Event Description
According to available information, this device required replacement due to a balloon burst.Three devices burst and fell out after placement and balloon test.One device had its balloon inflated with 45 ml of sterile water.Another balloon burst during the test with 20 ml of sterile water.The patient presented with hematuria after repeated attempts at placement.A device was successfully placed without any problem.
 
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-FLOW CATHETER
Type of Device
CATHETER, RETENTION TYPE, BALLOON
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebæk 3050
DA  3050
Manufacturer (Section G)
COLOPLAST A/S MANUFACTURING
9 avenue edmond rostand
sarlat-la-caneda 24206
FR   24206
Manufacturer Contact
usbes brian e schmidt
1601 west river road north
minneapolis, MN 55411
8007880293
MDR Report Key15670340
MDR Text Key304887925
Report Number9610711-2022-00087
Device Sequence Number1
Product Code EZL
UDI-Device Identifier03600040144442
UDI-Public03600040144442
Combination Product (y/n)N
PMA/PMN Number
K013172
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAB63221002
Device Catalogue NumberAB6322
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/30/2022
Initial Date FDA Received10/25/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-