Model Number CYF-V2 |
Device Problem
Material Rupture (1546)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/06/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported the bending section cover was ruptured off the distal end of the subject device.The reported issue was observed while reprocessing the subject device.There was no patient involvement.
|
|
Manufacturer Narrative
|
The subject device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, the reported issue (bending section cover ruptured) was confirmed.In addition, service found there was a leakage from the bending section cover and it was not possible to seal the bending section cover.The investigation is ongoing; therefore, a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the likely cause of the reported event is due to stress, handling, or other issues.However, the root cause of the reported event is unable to be determined.Olympus will continue to monitor field performance for this device.
|
|