MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. VIDEOSCOPE "CYF-V2", UK VERSION; CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. VIDEOSCOPE "CYF-V2", UK VERSION; CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-V2
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2022
Event Type  malfunction  
Event Description
The customer reported the bending section cover was ruptured off the distal end of the subject device.The reported issue was observed while reprocessing the subject device.There was no patient involvement.
 
Manufacturer Narrative
The subject device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, the reported issue (bending section cover ruptured) was confirmed.In addition, service found there was a leakage from the bending section cover and it was not possible to seal the bending section cover.The investigation is ongoing; therefore, a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the likely cause of the reported event is due to stress, handling, or other issues.However, the root cause of the reported event is unable to be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOSCOPE "CYF-V2", UK VERSION
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita,
aizuwakamatsu-shi,, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15670915
MDR Text Key306896308
Report Number9610595-2022-03320
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339431
UDI-Public04953170339431
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/07/2022
Initial Date FDA Received10/25/2022
Supplement Dates Manufacturer Received11/15/2022
Supplement Dates FDA Received12/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/02/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-