Model Number ALLURA XPER FD10
Device Problem Radiation Underexposure (3018)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2022
Event Type  malfunction  
Event Description
It was reported to philips that the allura device lost live imaging during a case.The system required multiple reboots in order for the procedure to be completed.No patient harm was reported.Philips has started an investigation for this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.The customer reported that the problem case x-ray output was lost.The system was in clinical use at the time of issue occurred.The customer had rebooted the system, it started to continue working.But there is no response from the distributor end.After reboot the system continued to work.The codes were updated based on the investigation outcome.Health impact code was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15671912
MDR Text Key303569113
Report Number3003768277-2022-00652
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGG
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/23/2022
Initial Date FDA Received10/26/2022
Supplement Dates Manufacturer Received12/18/2023
Supplement Dates FDA Received12/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-