MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN CHS COMPRESSION HIP SCREW; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN CHS COMPRESSION HIP SCREW; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Non-union Bone Fracture (2369)
Event Date 05/01/2000
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).Doi: 10.1097/01241398-200005000-00013.
 
Event Description
It was reported that, on the literature review "which implant for proximal femoral osteotomy in children?.A comparison of the ao (asif) 90[degrees] fixed-angle blade plate and the richards intermediate hip screw", one (1) patient suffered from a painful non-union, after a primary osteotomy was performed, using the compression hip screw system.The fixation system was revised with blade plate fixation 6 months after.The patient suffered from a deep infection after that, but the outcome is unknown.No further information is available.
 
Manufacturer Narrative
H3, h6: given the nature of the alleged incident, the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, as of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to report painful non-union.Per the literature review, no further information is available.The images provided in the article have been interpreted within the text; therefore, no further analysis of the images is required.The impact to the patient beyond the reported painful non-union, the revision and the deep infection cannot be confirmed no concluded due to the limited information provided.Should any additional relevant medical information be provided, this case would be re-assessed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, excessive pressure on the joint, procedural/user error, surgical/post operative complications, healing issues and/or patient condition.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN CHS COMPRESSION HIP SCREW
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15674695
MDR Text Key302435374
Report Number1020279-2022-04568
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Literature
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/03/2022
Initial Date FDA Received10/26/2022
Supplement Dates Manufacturer Received11/14/2022
Supplement Dates FDA Received11/18/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age15 YR
Patient SexFemale
-
-