• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM MEXICO FLAT DECK 36" WIDE WITH X-RAY; BED, FLOTATION THERAPY, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM MEXICO FLAT DECK 36" WIDE WITH X-RAY; BED, FLOTATION THERAPY, POWERED Back to Search Results
Model Number P7924A01
Device Problems Inflation Problem (1310); Component or Accessory Incompatibility (2897)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 10/05/2022
Event Type  Injury  
Manufacturer Narrative
The customer reported that the pro+ mattress was compressed and that an increase in the development of pressure injuries was associated with its use.The customer report could not associate a patient with a specific bed/serial number but did provide a room number the patient occupied during the hospitalization.The reported injuries associated with patients 1- 7 have been evaluated separately.This evaluation addresses patient 2.The patient was admitted to the hospital for severe sepsis.The patient did not have a pre-existing pressure ulcer at the time of admission; however, the customer reported that this patient developed bilateral deep tissue injury to right and left buttock.Multiple attempts to obtain additional details of the reported injury including medical intervention, past medical history, accessory devices utilized, and facilities positioning protocols were unsuccessful.The customer declined to provide any additional details related to the patient or the injury.The pro+ mattress is intended for patient support and for the prevention and/or treatment of pressure injuries.The pro+ mattress is intended for use in healthcare environments and is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.The mattress supports patients weighing between 70 lb (32 kg) and 500 lb (227 kg).The ifu notes: the pro+ mattress is not a substitute for good nursing practices.The mattress should be used in conjunction with good assessment and protocol.Additionally, the ifu instructs "do not use a mattress that has a damaged cover or mattress core assembly.If a deformity is on the mattress, determine the amount of deformity by laying a straight edge across the mattress and measuring the distance of the largest gap in the patient area.Note: some deformation is normal as mattresses age.¿ determine if the deformation is excessive per facility standards.Replace or repair as necessary." the development of pressure injuries is multifactorial and cannot only be attributed to the performance of the surface or the pro plus mattress.Pressure injuries can develop within 2-6 hours when capillaries supplying the skin and subcutaneous tissues are compressed, causing an obstruction in blood flow, which ultimately leads to tissue necrosis.Position changes are key to pressure injury prevention and treatment.These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas.Prevention of pressure injuries involves a multidisciplinary approach including rapid identification of at-risk individuals such as those with diabetes, incontinence, impaired mobility, peripheral vascular disease, etc.And exercising a vigilant prevention protocol consisting of skincare, nutritional assessment reducing mechanical load, and utilizing support surfaces.(dtpi) are persistent non-blanchable deep red, purple or maroon areas of intact skin, non-intact skin or blood-filled blisters caused by damage to the underlying soft tissues.Ulcers covered with slough or eschar are unstageable.Treatment can include frequent repositioning off the site of injury, good skin care, proper support surface selection, correcting any systemic issues or nutritional deficiencies and medical treatment from a wound care specialist.At times additional treatments can include prescribed antibiotic therapy and removal of dead tissue to promote healing and prevent or treat infection.In some cases, a dti could resolve in few days, becoming pi stage 1 or 2 once they appear on the external layers of the skin.In other cases, they could develop into stage 3 or 4 (or unstageable) depending on the involvement of deeper underlying structure.Based on the limited details available, the root cause of the reported event cannot be determined and the severity of this being a critical/serious injury cannot be excluded at this time.Additionally, at this time functionality of the device cannot be confirmed as this investigation is still ongoing to determine specific details of the reported event.If any additional relevant details are received, the complaint will be addressed accordingly.No further information is available on the repair of the bed at this time.The investigation is ongoing, however if any additional relevant information is identified following completion of the investigation, the additional relevant information will be submitted in a supplemental report.
 
Event Description
The customer reported that the pro+ mattress was compressed and that an increase in the development of pressure injuries was associated with its use.The customer report could not associate a patient with a specific bed/serial number but did provide a room number the patient occupied during the hospitalization.The reported injuries associated with patients 1- 7 have been evaluated separately.This evaluation addresses patient 2.This report was filed in our complaint handling system as complaint #(b)(4).
 
Event Description
The customer reported that the pro+ mattress was compressed and that an increase in the development of pressure injuries was associated with its use.The customer report could not associate a patient with a specific bed/serial number but did provide a room number the patient occupied during the hospitalization.The reported injuries associated with patients 1- 7 have been evaluated separately.This evaluation addresses patient 2.This report was filed in our complaint handling system as complaint #(b)(6).
 
Manufacturer Narrative
The customer reported that the pro+ mattress was compressed and that an increase in the development of pressure injuries was associated with its use.The customer report could not associate a patient with a specific bed/serial number but did provide a room number the patient occupied during the hospitalization.The reported injuries associated with patients 1- 7 have been evaluated separately.This evaluation addresses patient 2.The patient was admitted to the hospital for severe sepsis.The patient did not have a pre-existing pressure ulcer at the time of admission; however, the customer reported that this patient developed bilateral deep tissue injury to right and left buttock.Multiple attempts to obtain additional details of the reported injury including medical intervention, past medical history, accessory devices utilized, and facilities positioning protocols were unsuccessful.The customer declined to provide any additional details related to the patient or the injury.The pro+ mattress is intended for patient support and for the prevention and/or treatment of pressure injuries.The pro+ mattress is intended for use in healthcare environments and is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.The mattress supports patients weighing between 70 lb (32 kg) and 500 lb (227 kg).The ifu notes: the pro+ mattress is not a substitute for good nursing practices.The mattress should be used in conjunction with good assessment and protocol.Additionally, the ifu instructs "do not use a mattress that has a damaged cover or mattress core assembly.If a deformity is on the mattress, determine the amount of deformity by laying a straight edge across the mattress and measuring the distance of the largest gap in the patient area.Note: some deformation is normal as mattresses age.Determine if the deformation is excessive per facility standards.Replace or repair as necessary." the development of pressure injuries is multifactorial and cannot only be attributed to the performance of the surface or the pro plus mattress.Pressure injuries can develop within 2-6 hours when capillaries supplying the skin and subcutaneous tissues are compressed, causing an obstruction in blood flow, which ultimately leads to tissue necrosis.Position changes are key to pressure injury prevention and treatment.These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas.Prevention of pressure injuries involves a multidisciplinary approach including rapid identification of at-risk individuals such as those with diabetes, incontinence, impaired mobility, peripheral vascular disease, etc.And exercising a vigilant prevention protocol consisting of skincare, nutritional assessment reducing mechanical load, and utilizing support surfaces.(dtpi) are persistent non-blanchable deep red, purple or maroon areas of intact skin, non-intact skin or blood-filled blisters caused by damage to the underlying soft tissues.Ulcers covered with slough or eschar are unstageable.Treatment can include frequent repositioning off the site of injury, good skin care, proper support surface selection, correcting any systemic issues or nutritional deficiencies and medical treatment from a wound care specialist.At times additional treatments can include prescribed antibiotic therapy and removal of dead tissue to promote healing and prevent or treat infection.In some cases, a dti could resolve in few days, becoming pi stage 1 or 2 once they appear on the external layers of the skin.In other cases, they could develop into stage 3 or 4 (or unstageable) depending on the involvement of deeper underlying structure.The root cause of the reported event was the use of an incompatible frame with the mattress causing the material to deteriorate faster.However, a deep tissue pressure injury typically necessitates medical intervention to preclude permanent impairment, thus indicating a serious injury occurred with this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLAT DECK 36" WIDE WITH X-RAY
Type of Device
BED, FLOTATION THERAPY, POWERED
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
kayla miller
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key15676853
MDR Text Key302450709
Report Number3006697241-2022-00068
Device Sequence Number1
Product Code IOQ
UDI-Device Identifier00887761977907
UDI-Public010088776197790711210709
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model NumberP7924A01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/05/2022
Initial Date FDA Received10/26/2022
Supplement Dates Manufacturer Received10/05/2022
Supplement Dates FDA Received05/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-