A healthcare facility in vastras reported, via a fisher & paykel healthcare (f&p) field representative, that a mr290vx vented autofeed humidification chamber was found to be leaking after one day of use.It was discovered the following day that the patient had dried blood in their nose.Inspection of the chamber identified a leak at the connection of the chamber dome and base.No further patient consequence was reported.
|
A healthcare facility in sweden reported, via a fisher & paykel healthcare (f&p) field representative, that a mr290vx vented autofeed humidification chamber was found to be leaking after one day of use.It was discovered the following day that the patient had dried blood in their nose.Inspection of the chamber identified a leak at the connection of the chamber dome and base.No further patient consequence was reported.
|
(b)(4).Method: the complaint mr290vx vented autofeed humidification chamber was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is therefore based on the information provided by the customer, previous investigations of similar complaints and our knowledge of the product.Results: the customer reported that a mr290vx vented autofeed humidification chamber was leaking during patient use.It was reported that the leak was between the chamber dome and base and that the customer had used nebulised drugs with sodium chloride with the complaint chamber.Conclusion: without the return of the complaint device, we are unable to determine the cause of the reported event.However, it is likely due to the use of sodium chloride.Sodium chloride is highly corrosive to aluminium.The presence of this solution could cause degradation of the aluminium plate over time.Every mr290vx chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290vx chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290vx vented autofeed humidification chamber state the following: "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions could lead to a loss of ventilation pressure.".
|