Model Number M3536A |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/02/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
|
Event Description
|
In the process of rescuing the patient, the doctor found that the electrode plate could not be hung after rescuing the patient, and after inspection, it was found that the hanging plates on both sides were cracked.
|
|
Manufacturer Narrative
|
Available details indicate that the device was evaluated by the hospital staff during a self-test, and not when the device was in use on a patient as initially reported.There was no patient harm.The hospital staff fastened the cracks on both sides of the hanging board with screws and were able to hang the paddles normally.No replacement component was needed.The customer repaired the cracked hanging assembly to resolve the issue.It has been concluded that no further action is required at this time.
|
|