MEDTRONIC, INC EXPORT AP ASPIRATION CATHETER; CATHETER, EMBOLECTOMY Back to Search Results
Catalog Number EXPORTAPCE
Device Problems Suction Problem (2170); Detachment of Device or Device Component (2907); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2022
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned.¿ these words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
 
Event Description
An attempt was made to use one 6f export advance aspiration catheter.The device was inspected with no issues noted.The device was prepped per ifu with no issues noted.It was reported that during a thrombectomy the device broke after the first aspiration.A second aspiration was not possible after flushing.No patient injury was reported.
 
Manufacturer Narrative
Additional information: it was later confirmed that the device just failed to aspire.It did not fracture into two parts or become deformed.The device successfully performed the first aspiration.The catheter was flushed immediately afterwards.On the second attempt at aspiration, it was not possible to aspirate.A new export catheter was used instead to aspirate the thrombus.The device did not break while in the patient.No intervention was required to remove the device.Initial reporter name and phone number provided.Additional fdd code added.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Product analysis: one 6f export ap aspiration catheter was received for analysis.A kink was noted approx.26cm distal to the strain relief.Deformation was evident to the proximal end of the guidewire lumen.Approx.2mm of guidewire lumen had lifted from the shaft.It was not possible to load a 0.014 inch guidewire through the guidewire lumen due to the deformation noted.Using a pin gauge to block the aspiration lumen the device was pressurised, however there was no evidence of a leak on the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPORT AP ASPIRATION CATHETER
Type of Device
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
Manufacturer (Section G)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15689783
MDR Text Key303653148
Report Number1220452-2022-00081
Device Sequence Number1
Product Code DXE
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K081573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/13/2023
Device Catalogue NumberEXPORTAPCE
Device Lot Number0011021069
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2022
Initial Date FDA Received10/28/2022
Supplement Dates Manufacturer Received11/14/2022
12/22/2022
Supplement Dates FDA Received11/28/2022
01/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-