Model Number CYF-VA2
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The olympus representative reported on behalf of the customer during cleaning, the visera cysto-nephro videoscope presented with an air leak on the distal end.No patient injury or procedure impact was reported.During the evaluation of the device, it was noted that there was scraping on the forceps plug cap.This report is to capture the reportable malfunction of the scraping on the forceps plug cap noted at estimation.
 
Manufacturer Narrative
Date of event/event: no additional information was available.The device was evaluated by olympus.The evaluation found the air leak on the distal end was confirmed.The forceps plug cap had been scraped off.The water tightness was not maintained due to holes in the bending section and perforation in the forceps channel.Wrinkles were found in the insertion tube and the bending angle was insufficient due to the elongation of the angle wire.The universal cord was collapsed, the video cable and the flexible tube of the insertion section were crushed, and the video connector case was cracked due to external factors.Scratches were found in multiple locations on the device.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to stress or handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15690767
MDR Text Key306897496
Report Number3002808148-2022-03637
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339455
UDI-Public04953170339455
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VA2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/11/2022
Initial Date FDA Received10/28/2022
Supplement Dates Manufacturer Received11/15/2022
Supplement Dates FDA Received12/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/29/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-