Updated fields: d4, d9, g3, g6, h2, h3, h6, h10.The duo headlight 2 bay system (90620us) was returned for evaluation.The duo headlight 2 bay system was received in used condition with cooling fan that was cutting on and off.Evaluation was able to duplicate the reported problem with the unit.The issue of the cooling fan for this duo headlight cutting on and off is most likely due to damage to the harness wire and fan caused by rough handling/environmental damage.The harness wire and fan were replaced to correct this issue.An evaluation and function test were performed and the unit passed all tests.The reported complaint was confirmed.No manufacturing, workmanship, or material deficiency has been identified.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|