• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO BRONCHOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO BRONCHOSCOPE Back to Search Results
Model Number EB-1970K
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2022
Event Type  malfunction  
Event Description
The time of event is unknown.There was no report of patient harm.Operation/suction channel clogged.
 
Manufacturer Narrative
We checked the returned unit and confirmed that the operation channel clogged.Based on the result, we concluded that it was caused due to the insufficient reprocessing at the facility.In addition, we confirmed that the light guide cable buckled, and the insertion flexible tube (ift) stuck accessory on the exterior; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report"hr-rpt-0588(channel)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO BRONCHOSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key15697681
MDR Text Key307028664
Report Number9610877-2022-59365
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K131028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEB-1970K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/11/2022
Initial Date FDA Received10/30/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-