ZIMMER BIOMET, INC. 36MM COCR MOD HD -3MM; PROSTHESIS, HIP Back to Search Results
Catalog Number 11-363661
Device Problems Material Erosion (1214); Naturally Worn (2988)
Patient Problems Failure of Implant (1924); Metal Related Pathology (4530)
Event Date 10/11/2022
Event Type  Injury  
Event Description
It was reported that a patient underwent a hip revision approximately 11 years post implantation due to the head wearing into the liner causing metallosis.Attempts have been made and no further information is available.
 
Manufacturer Narrative
(b)(4).Concomitant medical products: unknown stem, cat #:131150 - mh solid shell ha/pc 50mm ln23 - lot #: 2423032, unknown liner.Foreign source: australia.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-02441.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, as supplemental medwatch will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6: component code: mechanical (g04) - head: visual examination of the provided pictures identified the liner has damage to the outer articulation surface.No wear can be seen on the inner surface in the provided picture.All components covered in bio-debris; no additional details can be seen on the provided images.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Devices are used for treatment.The reported products were reviewed for compatibility with no issues noted.Review of complaint history found no additional related issues for this/these item(s) and the reported part and lot combination(s).Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: there is abnormal superolateral subluxation of the femoral head within the cup.Bone quality is osteopenic.There is marked liner wear with malalignment of the femoral head within the cup as noted.Radiolucency is present along the proximal femur consistent with osteolysis.The malalignment/subluxation is secondary to marked liner wear.While metallosis is not seen radiographically, this would contribute to the osteolysis noted.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no update to the original complaint description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
36MM COCR MOD HD -3MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15698615
MDR Text Key302689372
Report Number0001825034-2022-02442
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K032396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2015
Device Catalogue Number11-363661
Device Lot Number506070
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/05/2022
Initial Date FDA Received10/31/2022
Supplement Dates Manufacturer Received11/21/2022
Supplement Dates FDA Received11/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
-
-