Model Number ALLURA FC
Device Problem Radiation Output Failure (4027)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2022
Event Type  malfunction  
Event Description
It was reported to philips that the exposure was not possible on the cardio vascular system-allura fc device.It is unknown if the device was in use at the time of the event as this information was not provided.No patent injury or harm was alleged.An investigation has been opened to evaluate this event.
 
Manufacturer Narrative
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite and confirmed that the system¿s ¿exposure was not possible properly¿ upon investigation, fse noticed the error¿s displayed on the system.As fse tried replacing ccb board, still the issue persists.As fse replaced frame rate board to solve the issue.After replacing generator touch screen console, to resolve touch sense issue, the system was returned to use in good working order, with required calibration done.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA FC
Type of Device
SYSTEM, X-RAY, FLUOROSCOPIC, IMAGE-INTENSIFIED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15704314
MDR Text Key304702235
Report Number3010685285-2022-00004
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K022899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA FC
Device Catalogue Number723003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/01/2022
Initial Date FDA Received10/31/2022
Supplement Dates Manufacturer Received09/04/2023
Supplement Dates FDA Received09/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-