Model Number 20E |
Device Problem
Incorrect Interpretation of Signal (1543)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/05/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Stryker will not request any patient identifying information to be in accordance with regulation (eu) 2016/679 of the european parliament and of the council.Stryker evaluated the customer's device and was unable to duplicate the reported issue.Proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer for use.
|
|
Event Description
|
A customer contacted stryker to report that their device detected a patient rhythm to be non-shockable, however the customer believed it to be shockable.As a result, wrong defibrillation therapy would be delivered, if needed.There were no reports of adverse effects to the patient as a result of the reported event.
|
|
Manufacturer Narrative
|
A stryker customer quality engineer evaluated the downloaded electronic patient records and was unable to verify the reported issue.There was no indication that the device was in automated external defibrillator (aed) mode.During the reported event, the device was in manual mode only where the device successfully delivered a shock.
|
|
Event Description
|
A customer contacted stryker to report that their device detected a patient rhythm to be non-shockable, however the customer believed it to be shockable.As a result, wrong defibrillation therapy would be delivered, if needed.There were no reports of adverse effects to the patient as a result of the reported event.
|
|
Search Alerts/Recalls
|