AIZU OLYMPUS CO., LTD. EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-H190
Device Problems Device Reprocessing Problem (1091); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was sent at regional repair center olympus france (rrc-ofr) for inspection and repair.The device evaluation still in progress and final evaluation report has not yet been finalized.Supplemental report(s) will be submitted should any relevant new information is available.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event Description
Customer reported with an issue of "air/water flow issues from the air/water flow system".The issue found during reprocessing.There was no patient harm, no user injury reported due to the event.Device return initial evaluation found clogging nozzle, foreign material cannot be removed due to buckling of the nozzle.This reported is being submitted for foreign material found and cannot be removed from the nozzle due to buckling.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the foreign material in the nozzle could not be determined, however, based on information provided by the customer, the issue is likely the result of the reprocessing deviating from the instructions for use (ifu).In addition, the following non-reportable malfunctions were found during the device evaluation: - a-rubber glue is worn out and separated - one lg-rod lens unit is cracked - air/water flow issues from the air/water flow system olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15705816
MDR Text Key307603644
Report Number9610595-2022-03533
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04953170305290
UDI-Public04953170305290
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-H190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/03/2022
Initial Date FDA Received11/01/2022
Supplement Dates Manufacturer Received02/21/2023
Supplement Dates FDA Received03/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-