Model Number LTF-S190-10
Device Problems Image Display Error/Artifact (1304); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
No additional information was available.The device was evaluated by olympus.The evaluation found that the allegation was confirmed.The scope presented a b31 scope communication error.An additional reportable malfunction was identified.The information identification function was lost due to the breakage of the identification board.Additional issues were identified.The hard tube was discolored, the curved pipe blade was broken, and the bending section adhesive part was missing.The vent metal on the light guide connector was loose scratches were found in multiple locations on the device.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The olympus representative reported the endoeye flex deflectable videoscope was displaying a b31 scope communication error code.The same scope communication error code was displayed even when the scope was connected to other systems.No patient injury or procedure impact was reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the events were unable to be identified; however, the events most likely occurred due to a failure of the image sensor unit or a problem with the circuit board inside the video connector.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15707994
MDR Text Key307278647
Report Number9610595-2022-03537
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170310386
UDI-Public04953170310386
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/06/2022
Initial Date FDA Received11/01/2022
Supplement Dates Manufacturer Received12/15/2022
Supplement Dates FDA Received01/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-