The customer reported incorrect monitoring results found during an operation, delaying treatment time.At 16:07 in the afternoon, the vital signs of patients were detected after coronary stent implantation.At 16:21, the equipment gave an alarm and found that the blood pressure and heartbeat were too low.The nurses found that the heartbeat was normal through manual measurement.After resetting the monitor, the detection was still inaccurate.The device was in clinical use a the time the issue was discovered.The reporter claimed potential injury of patient.
|
The customer reported incorrect monitoring results found during an operation, delaying treatment time.At 16:07 in the afternoon, the vital signs of patients were detected after coronary stent implantation.At 16:21, the equipment gave an alarm and found that the blood pressure and heartbeat were too low.The nurses found that the heartbeat was normal through manual measurement.After resetting the monitor, the detection was still inaccurate.The device was in clinical use a the time the issue was discovered.The reporter claimed potential injury of patient.The report was sent for technical investigation.The reported problem was not confirmed; the reported problem was unable to be reproduced.After replacing the ecg electrodes and blood pressure cuffs, the equipment returned to perform measurements normally.The investigation confirmed there was no harm to the patient; at the time of the reported event, the nurses verified that the heartbeat was normal through manual measurement.No further information was provided.
|