ZIMMER BIOMET, INC. M2A-MAGNUM PF CUP 56ODX50ID; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Patient Device Interaction Problem (4001)
Patient Problems Pain (1994); Metal Related Pathology (4530)
Event Date 04/26/2021
Event Type  Injury  
Event Description
It was reported that the patient underwent a right hip revision procedure approximately 9 years post-implantation.The patient suffered symptoms including but not limited to increasingly debilitating pain, discomfort, and soreness, permanent harm by severe metal poisoning and metallosis from the metal debris of the magnum device.No additional information.
 
Manufacturer Narrative
(b)(4).Concomitant medical products: unk stem; unk taper; unk head.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2022 - 02471; 0001825034 - 2022 - 02472; 0001825034 - 2022 - 02473.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: a2; a3; b4; b5; b7; g3; d1; d2; d4; g4; h2; h3; h4; h6 corrected d6b d6b ¿ upon the receipt and review of op notes, the cup remained implanted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that a patient underwent a right hip revision approximately nine years¿ post-implantation due to pain, elevated metal ions, metallosis, and metal debris.During the revision, it was noted there was wear to the trunnion.The cup and stem remained implanted with the head and liner exchange without complications.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Dhr was reviewed and no discrepancies related to the reported event were found.Medical records/radiographs were provided and reviewed by a health care professional.A review of the available records identified findings of the reported issues.Increasingly debilitating pain, discomfort, and soreness, permanent harm by severe metal poisoning and metallosis from the metal debris¿.Complaint is confirmed based on the provided medical records.Root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M2A-MAGNUM PF CUP 56ODX50ID
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15713496
MDR Text Key302827496
Report Number0001825034-2022-02470
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101336
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model NumberN/A
Device Catalogue NumberUS157856
Device Lot Number695380
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/07/2022
Initial Date FDA Received11/02/2022
Supplement Dates Manufacturer Received06/09/2023
09/19/2023
Supplement Dates FDA Received07/06/2023
09/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexMale
-
-