• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH Back to Search Results
Model Number G407208
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Perforation (2513)
Event Date 10/24/2022
Event Type  Injury  
Event Description
Following an atrial fibrillation procedure, an effusion was noted in the right atrium requiring a pericardiocentesis.During the post assessment, an ultrasound revealed a pericardial effusion.A pericardiocentesis was performed to stabilize the patient.There were no performance issues with any abbott devices.The physician indicated this occurred during transseptal puncture.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported perforation/effusion remains unknown.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TRANSSEPTAL NEEDLE
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15714736
MDR Text Key302848864
Report Number3008452825-2022-00659
Device Sequence Number1
Product Code DRC
UDI-Device Identifier05414734205139
UDI-Public05414734205139
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K122587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberG407208
Device Catalogue NumberG407208
Device Lot Number8287445
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/24/2022
Initial Date FDA Received11/02/2022
Supplement Dates Manufacturer Received01/17/2023
Supplement Dates FDA Received01/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-