PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10/10; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD10/10
Device Problems No Device Output (1435); Radiation Output Failure (4027)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2022
Event Type  malfunction  
Event Description
It has been reported to philips that on (b)(6) 2022 the system failed.There was a patient with a heart attack on the table.The patient had to be moved to another room.No harm has been reported to philips.To date, no further information was received.Philips has started an investigation and a follow up report will be submitted when further information is received.
 
Manufacturer Narrative
Philips has investigated this complaint.There was no additional information regarding if the patient's condition (mi)worsened during the transfer/delay and if the patient required any additional treatment during the transfer.The philips remote service engineer (rse) analyzed the system remotely and confirmed the reported issue.Upon troubleshooting, the rse determined further grid switch errors and tube errors reported in the logging.The customer had been quoted for a new tube.However the service quote provided by philips was not accepted by the customer, therefore no further action could be performed by philips.The codes were updated based on the investigation outcome.Device problem code was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10/10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15715932
MDR Text Key304886015
Report Number3003768277-2022-00731
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10/10
Device Catalogue Number722005
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/21/2022
Initial Date FDA Received11/02/2022
Supplement Dates Manufacturer Received10/03/2023
Supplement Dates FDA Received10/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-