• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE ENVELLA BED W/TRAPEZE; BED, AIR FLUIDIZED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE ENVELLA BED W/TRAPEZE; BED, AIR FLUIDIZED Back to Search Results
Model Number P0821A
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/10/2022
Event Type  malfunction  
Event Description
Hillrom received a report from a hillrom technician stating the bed cord had an electric spark, burning smell, and did not function.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # : (b)(4).
 
Manufacturer Narrative
The hillrom technician found the pe inlet and cord needed to be replaced.The envella® air fluidized therapy system is intended for medical purposes to help treat or prevent pressure injuries, to treat severe or extensive burns, or to aid in circulation.This bed is an ideal support for patients who have advanced pressure injuries, flaps, grafts, or burns, and require frequent transfers or variable head elevation, and any other conditions appropriate for air fluidized therapy.The bed also can be used for intractable pain, extensive epidermal detachment, stevens-johnson syndrome, purpura fulminans, induce relaxation which may reduce need for sedative and pain medication, improve patient outcome, and reduce wound healing time.The bed permits easy positioning and egress, thereby enhancing the independence of patients.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in february 2022.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the pe inlet and cord to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVELLA BED W/TRAPEZE
Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
brad wheeler
1069 state route 46 east
batesville, IN 47006
3128199307
MDR Report Key15718935
MDR Text Key303317182
Report Number1824206-2022-00481
Device Sequence Number1
Product Code INX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP0821A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/10/2022
Initial Date FDA Received11/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/17/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-