• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR Back to Search Results
Model Number M3535A
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2022
Event Type  malfunction  
Event Description
It was reported to philips that the paddle tray is cracked.There was no patient involvement.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Manufacturer Narrative
Philips received a complaint on the heartstart mrx monitor/defib indicating that the paddle tray was cracked.There was reportedly no patient involvement.A philips field service engineer (fse) evaluated the device and confirmed that there was physical damage to the paddle tray.The fse replaced the paddle tray to resolve the issue.The device passed all performance assurance tests and was returned to service.The device remains at the customer's site.No further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX MONITOR/DEFIB
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
sara carlson
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key15721753
MDR Text Key305063386
Report Number3030677-2022-04950
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838000018
UDI-Public00884838000018
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3535A
Device Catalogue NumberM3535A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/07/2022
Initial Date FDA Received11/03/2022
Supplement Dates Manufacturer Received10/07/2022
Supplement Dates FDA Received03/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-