Model Number V60
Device Problem Blocked Connection (2888)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2022
Event Type  malfunction  
Event Description
Philips received a complaint from the customer on the v60 indicating that a patient disconnect-patient circuit occluded error occurred.It was reported there was no patient involvement at the time the issue was discovered.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The customer reported that no air was blowing from the patient outlet.Customer went to the pneumatics page in diagnostics and set a flow of 50slpm.Blower voltage did not increase, and blower rpm stayed at 3000rpm.Followed troubleshooting of 1134 (blower stalled) code as per service manual.The rse instructed the customer based on the steps within the service manual.The rse recommended replacement of the blower assembly and provided part number.The investigation is ongoing.
 
Manufacturer Narrative
The customer replaced blower assembly to resolve the reported issue.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
kimberly shelly
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key15725363
MDR Text Key307306385
Report Number2031642-2022-02720
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/11/2022
Initial Date FDA Received11/04/2022
Supplement Dates Manufacturer Received11/07/2022
Supplement Dates FDA Received11/20/2022
Date Device Manufactured12/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-