Model Number V60
Device Problem Failure to Read Input Signal (1581)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2022
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating that the unit would not stay in standby mode.The device was being used on a patient at the time of the failure; there was no patient or user harm reported.The remote service engineer (rse), while troubleshooting, noted that the average air was reading -31.2 slpm.The remote service engineer (rse) recommended replacing the flow sensor assembly and provided the part number.The customer will order the flow sensor assembly and repair the unit.The investigation is ongoing.
 
Manufacturer Narrative
Insufficient information is available to determine the resolution of the event.The customer was provided with parts id; however, it could not be confirmed if the customer replaced the specified parts.Multiple good faith efforts (gfe) were performed for further details regarding the event; however, no response was provided.No parts were returned for failure investigation.In the event that parts are returned or if new information is provided, the complaint will be reopened to update the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
kimberly shelly
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key15725835
MDR Text Key306981491
Report Number2031642-2022-02731
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/11/2022
Initial Date FDA Received11/04/2022
Supplement Dates Manufacturer Received01/04/2023
Supplement Dates FDA Received01/11/2023
Date Device Manufactured11/29/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-