RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Read Input Signal (1581)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/11/2022 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint on the v60 ventilator indicating that the unit would not stay in standby mode.The device was being used on a patient at the time of the failure; there was no patient or user harm reported.The remote service engineer (rse), while troubleshooting, noted that the average air was reading -31.2 slpm.The remote service engineer (rse) recommended replacing the flow sensor assembly and provided the part number.The customer will order the flow sensor assembly and repair the unit.The investigation is ongoing.
|
|
Manufacturer Narrative
|
Insufficient information is available to determine the resolution of the event.The customer was provided with parts id; however, it could not be confirmed if the customer replaced the specified parts.Multiple good faith efforts (gfe) were performed for further details regarding the event; however, no response was provided.No parts were returned for failure investigation.In the event that parts are returned or if new information is provided, the complaint will be reopened to update the investigation.
|
|