AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V3
Device Problem Separation Problem (4043)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2022
Event Type  malfunction  
Event Description
The customer reported to olympus that there was a water leak, while using the uretero-reno videoscope.The issue was found after the therapeutic procedure (f-tul), there was no delay and the procedure was completed with the same set of equipment.There was no patient harm associated with this event.The device was returned and evaluated, and it was found that the curved rubber adhesive had fallen off and was missing.This mdr is being submitted to capture the reportable malfunction found during the device evaluation.
 
Manufacturer Narrative
The device was returned and evaluated and the customers allegation was confirmed; watertightness was not maintained due to holes in the curved rubber.The curved rubber adhesive had fallen off (was missing), the bending angle was over, a cut on the angle wire was recognized, the forceps could not be inserted smoothly into the forceps channel, scratches were found on the insertion tube and scratches were found on the operation part.Scratches were also found on the following parts: the rotating mechanism, the grip, the universal cord, the video cable, the video connector, the video connector case, the light guide connector, and the up/down angle fixing lever.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely due to external factors or handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15729949
MDR Text Key307468415
Report Number9610595-2022-03604
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403385
UDI-Public04953170403385
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/10/2022
Initial Date FDA Received11/04/2022
Supplement Dates Manufacturer Received11/21/2022
Supplement Dates FDA Received12/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/21/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment