Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2022
Event Type  malfunction  
Event Description
It was reported stem was opened, and the packaging showed signs of wear.The surgeon decided not to implant.No additional information on the reported event.
 
Manufacturer Narrative
(b)(4).Evaluation of the returned product confirmed there is white debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier.There is debris inside the sterile packaging which is consistent with the appearance of porous coating.Identified damage to the sterile packaging (pouch).Sterility cannot be confirmed as the blister package was opened upon arrival.Review of the device history record(s) identified no deviations or anomalies during manufacturing.The likely condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR T1 PPS 18X156MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15732496
MDR Text Key304344970
Report Number0001825034-2022-02494
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00880304517196
UDI-Public(01)00880304517196(17)260302(10)3739614
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-105180
Device Lot Number3739614
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/03/2022
Initial Date FDA Received11/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberNI
Patient Sequence Number1
-
-