• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT PT/INR CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT PT/INR CARTRIDGE Back to Search Results
Catalog Number 03P89-24
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/13/2022
Event Type  malfunction  
Event Description
On (b)(6) 2022, abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded a suspected discrepant pt/inr result on a 75 year old male patient with deep vein thrombosis.There was no additional patient information available, at the time of this report.Return product is not available for investigation.Method: i-stat, date:(b)(6) 2022, result: >4.5.Method: acl top 750, date: (b)(6) 2022, results: 3.8.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Manufacturer Narrative
Apoc incident #: (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Manufacturer Narrative
Apoc incident: # (b)(4).The investigation was completed on 21-nov-2022.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Retained cartridge testing met the acceptance criteria in appendix 1 of q04.01.003 rev.Ak (product complaint level 2 and level 3 investigation procedure).Returned cartridge testing was unable to reproduce the customer's observation of unexpected results.Although the sample size was not sufficient to make a pass/fail determination for rate of results outside allowable error (ea), returned cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev ak.(product complaint level 2 and level 3 investigation procedure) with respect to rate of suppressed results.No deficiency has been identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT PT/INR CARTRIDGE
Type of Device
PT/INR CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key15734962
MDR Text Key307001208
Report Number2245578-2022-00153
Device Sequence Number1
Product Code GJS
UDI-Device Identifier10054749000170
UDI-Public10054749000170
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020355
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/21/2023
Device Catalogue Number03P89-24
Device Lot NumberT22206A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2022
Initial Date FDA Received11/04/2022
Supplement Dates Manufacturer Received11/21/2022
Supplement Dates FDA Received11/22/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexMale
-
-