Model Number ECY-1575K
Device Problem Optical Distortion (3000)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2022
Event Type  malfunction  
Manufacturer Narrative
This device is classified as import for export, therefore 510k is not applicable.Model eb-1575k is available in the usa with a 510k number k131028.We checked the returned unit and confirmed that the image cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the ccd module.In addition, we confirmed that the u/d knob broken, the insertion flexible tube attaching nut cracked, the control body corroded, the ccd driver pcb corroded, the electrical connector corroded, the lg cable connector corroded, the insertion flexible tube (ift) leak, the light guide cable bump, the operation channel resistance, the angulation-down angulation decrease, and the angulation-up angulation decrease; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event Description
The time of event is unknown.There was no report of patient harm.Video image failure(cloudy ).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO CYSTOSCOPE 2.0C 5.5TP 400L
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key15741792
MDR Text Key307505103
Report Number9610877-2022-59635
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberECY-1575K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2022
Initial Date FDA Received11/07/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-