MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
|
Back to Search Results |
|
Model Number 3058 |
Device Problems
Failure to Interrogate (1332); Inadequacy of Device Shape and/or Size (1583); Communication or Transmission Problem (2896); Impedance Problem (2950)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/03/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that nurse had measured impedance problems. due to wrong impedance measurement, the patient was in or for lead testing and battery change.In operating room (or) the ins could not be found.While explanting the ipg we saw that there was something inside the ins.It looks like erosion.The lead was inspected and was looking normal.With the new ins the impedance check was fine.
|
|
Event Description
|
Additional information was received.They reported the device has been sent for analysis.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: analysis of the ins (s/n (b)(6) revealed that there was no output or telemetry on the implantable neurostimulator (ins) and determined normal battery depletion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|