Model Number 29MECJ-502
Device Problem Perivalvular Leak (1457)
Patient Problem High Blood Pressure/ Hypertension (1908)
Event Date 10/13/2022
Event Type  Injury  
Event Description
It was reported that on (b)(6) 2022 a 29mm sjm masters series valve expanded cuff was implanted successfully.On an unknown date, the patient developed a severe paravalvular leak.It was noted that during the same time the patient had jaundice, unrelated to the paravalvular leak, that resolved after treatment.The valve was explanted (b)(6) 2022 and replaced with a new 29mm sjm masters series valve expanded cuff.It was noted that the patient experienced pulmonary hypertension during surgery.The patient is reported to be stable.No additional information was reported.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
Explant due to perivalvular leak was reported.The investigation found that both leaflets were intact and opened and closed completely without difficulty.No material was seen within the pivot recesses or on the surfaces of the mechanical leaflets.The sewing cuff was intact.There was a patchy fibrous pannus formation on the sewing cuff which did not appear to impinge upon the mechanical components.No inflammation.Sections of material removed from the sewing cuff consisted of fibrous tissue consistent with pannus.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received the cause of the reported incident could not conclusively be determined.H1: report type corrected to serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERS SERIES VALVE EXPANDED CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15750063
MDR Text Key307015437
Report Number2135147-2022-01970
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734006545
UDI-Public05414734006545
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number29MECJ-502
Device Catalogue Number29MECJ-502
Device Lot Number7856556
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/13/2022
Initial Date FDA Received11/07/2022
Supplement Dates Manufacturer Received02/01/2023
Supplement Dates FDA Received02/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexMale
-
-