Model Number CLV-190 |
Device Problems
Communication or Transmission Problem (2896); Incomplete or Inadequate Connection (4037)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/11/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
In speaking with olympus technical assistance center (tac), the customer reported receiving an error code b30 message with all scopes.It was noted the customer was not cycling power when swapping the scopes.The customer was asked to use an alcohol prep pad to clean the contacts on the scopes.A scope was inserted with the power off then the power was cycled.The clv-190 was retested without issue.The device wasn't returned because the issue was able to be troubleshooted and resolved.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
|
|
Event Description
|
The customer contacted the olympus technical assistance center (tac), to report the evis exera iii xenon light source had an error code b30 message with all scopes.There were no reports of patient harm associated with this event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Three attempts were performed to obtain additional information, but no response was received from the customer.The device was not returned; therefore, the reported phenomenon or condition of the device could not be confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It was confirmed that the device was shipped according to the specifications.It has been over 7 years since the subject device was manufactured.The specific root cause of the reported problem could not be determined at this time because the device was not returned.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|