SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problems Communication or Transmission Problem (2896); Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2022
Event Type  malfunction  
Manufacturer Narrative
In speaking with olympus technical assistance center (tac), the customer reported receiving an error code b30 message with all scopes.It was noted the customer was not cycling power when swapping the scopes.The customer was asked to use an alcohol prep pad to clean the contacts on the scopes.A scope was inserted with the power off then the power was cycled.The clv-190 was retested without issue.The device wasn't returned because the issue was able to be troubleshooted and resolved.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer contacted the olympus technical assistance center (tac), to report the evis exera iii xenon light source had an error code b30 message with all scopes.There were no reports of patient harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Three attempts were performed to obtain additional information, but no response was received from the customer.The device was not returned; therefore, the reported phenomenon or condition of the device could not be confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It was confirmed that the device was shipped according to the specifications.It has been over 7 years since the subject device was manufactured.The specific root cause of the reported problem could not be determined at this time because the device was not returned.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15752113
MDR Text Key307663295
Report Number3002808148-2022-03915
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/11/2022
Initial Date FDA Received11/08/2022
Supplement Dates Manufacturer Received12/12/2022
Supplement Dates FDA Received01/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/12/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse