The device was returned to zoll medical corporation.The customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, impedance testing, defibrillation cycling and environmental chamber testing without duplicating the.Report.Without the clinical file, the algorithm decision could not be evaluated based on the segment in question.The device was recertified and returned to the customer with no fault found.Analysis of reports of this type has not identified an increase in trend.
|