Reportable based on device analysis completed on (b)(6) 2022.It was reported that stent difficulty advancing and deployment difficulty occurred.The 75% stenosed target lesion was located in the severely tortuous iliac artery.A 8x80x75 epic vascular stent was advanced for treatment.However, during stent deployment, the thumbwheel was rotated, but marker on the outer shaft was not moving under fluoroscopy, and the stent could not be deployed.The rotation of the thumbwheel felt heavy, but it did not come off or break down.The resistance was stronger than usual because the meandering of the blood vessels was strong, and the placement of the stent was difficult.The use of the device was stopped, and it was removed from the sheath as it was.The procedure was completed with the balloon only and there were no patient complications nor injuries reported.However, returned device analysis revealed that the stent was partially deployed.
|
Initial reporter address1: (b)(6).Initial reporter city: (b)(6).Device evaluated by mfr.: returned product consisted of an epic self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the stent is partially deployed 4.8cm from the distal end of the middle sheath.The shaft is damaged 13.5cm from the distal end of the middle sheath.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.
|