• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS® EGFR MUTATION TEST V2; SOMATIC GENE MUTATION DETECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS® EGFR MUTATION TEST V2; SOMATIC GENE MUTATION DETECTION SYSTEM Back to Search Results
Catalog Number 07248563190
Device Problem False Positive Result (1227)
Patient Problems Solid Tumour (4552); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2022
Event Type  malfunction  
Manufacturer Narrative
Roche received complaints from customers reporting the generation of false mutation detected results for the exon 20 insertion (ex20ins) mutation when using the cobas® egfr mutation test v2.During in-house testing using customer-provided ffpet samples, an ex20ins false mutation detected result was reproduced for one out of 8 ffpet samples, which was processed following the validated sample preparation method from the instructions for use.Although the majority of cases reported were from users using ffpet samples, the generation of false mutation detected ex20ins results with plasma specimens was reported in one case.A false mutation detected ex20ins result could lead to harm under specific scenarios.Consignees have been notified of the issue with instruction to follow the instructions for use for sample input requirements.Additionally, if an ex20ins mutation detected result is generated with the cobas® egfr mutation test v2, customers must confirm the result with another method (e.G., sequencing or other pcr-based tests).E1 facility name truncated due to character limit : (b)(6).
 
Event Description
A customer from france alleged discrepant results with 1 patient sample using the cobas® egfr mutation test v2.The sample was initially tested using the cobas® egfr mutation test v2 and generated a "mutation detected" result for exon20ins.The same sample was retested using next generation sequencing (ngs) and ex20ins was not detected.The positive results was not reported.No harm was alleged.An investigation was conducted to evaluate the customer issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS® EGFR MUTATION TEST V2
Type of Device
SOMATIC GENE MUTATION DETECTION SYSTEM
Manufacturer (Section D)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer (Section G)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
na
branchburg NJ 08876
Manufacturer Contact
stacie-ann creighton
1080 us hwy 202 s
na
branchburg, NJ 08876
9082537112
MDR Report Key15756237
MDR Text Key307714178
Report Number2243471-2022-00916
Device Sequence Number1
Product Code OWD
UDI-Device Identifier00875197005448
UDI-Public00875197005448
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P150047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Catalogue Number07248563190
Device Lot NumberH31955
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/21/2022
Initial Date FDA Received11/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number2243471-08-24-2021-003-C
Patient Sequence Number1
-
-