Model Number 5711012
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2022
Event Type  malfunction  
Event Description
Olympus (oekg) was informed by quality assurance specialist at the user facility that the customer received the sur-catch helical basket which ¿appeared to have moulding defects during packing from the factory¿.The customer problem was found during a receipt inspection.No death, injury or harm was reported to olympus.
 
Manufacturer Narrative
The subject device was returned to the olympus repair center (oekg).Also, images were provided of the device in sealed packaging.The packaging of the device shows there is slight warping.The device history records for this product have been reviewed.All records indicate that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.A total of 12 units were produced under this lot number with no associated non-conformances, reported scrap or recorded process deviations relating to the reported failure.The investigation is ongoing.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it¿s likely the damaged packaging was due to transportation and/or storage at high temperatures.The root cause of this event was unable to be identified.The following is included in the instructions for use: ¿should the product or its container be received damaged, do not use.Sterile units are ready for use if the package is intact.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUR-CATCH HELICAL BASKET, 3.0FR,3PR,5CM TIP
Type of Device
DISLODGER, STONE, FLEXIBLE
Manufacturer (Section D)
GYRUS ACMI L.P.
9600 louisiana ave north
brooklyn park MN 55445
Manufacturer Contact
todd brill
9600 louisiana avenue north
brooklyn park, MN 55445
5082077661
MDR Report Key15756840
MDR Text Key307718814
Report Number3011050570-2022-00123
Device Sequence Number1
Product Code FGO
UDI-Device Identifier00821925010772
UDI-Public00821925010772
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
EXEMPT-FGO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5711012
Device Lot NumberKR243620
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/12/2022
Initial Date FDA Received11/08/2022
Supplement Dates Manufacturer Received11/21/2022
Supplement Dates FDA Received12/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-